Which technique listed below has the best DL for folic acid given the values for the DL shown here? Circle the correct answer.
HPLC, with a DL = 1.8 μg/mL;
Fluorescence, with a DL = 0.50 μg/mL;
Absorbance, with a DL= 2.6 μg/mL.
Ans. “Detection limit is the lowest concentration of the analyte that can be detected”.
The smaller the DL, the better.
# For example, an UV-vis spectrophotometer has a DL of 100.0 ug/mL for a hypothetical analyte.
HPLC has the DL of 1.8 ug/mL for the same analyte.
If a solution has [analyte] = 10.0 ug/ mL , the UV-vis spectrophotometer fails to detect the analyte in the sample. However, HPLC can detect it .
# Therefore, the smaller the DL, the better.
So, correct option is- Fluorescence, with a DL = 0.50 ug/ mL