Question

2) Accuracy In order to assess the accuracy of each technique a sample of a standard...

2) Accuracy

In order to assess the accuracy of each technique a sample of a standard reference material NIST SRM 3280 was analysed. The certificate of analysis is available on Blackboard for you to consult. The preparation of the standard is given below. Using the data provided check the recovery of As obtained for both techniques.

A 0.5094g sample of NIST 3280 was digested in 10ml of nitric acid (8ml) and hydrochloric acid (2ml). The resulting solution was diluted to 25ml. This was then analysed by ICP-OES (calibrated 0-20ng/ml) and ICP-MS (calibrated 0-10ng/ml). The resulting solution gave the following calibration data and signals for the solution being analysed

ICP- OES

ICP-MS

Equation pf Calibration Line

y=1.1713x + 0.3892

y=23.735x + 1.8857

Average Sample Reading (n-3)

4.064 +/- 0.50 cts

65.132 +/- 1.21 cts

Using the numbers above calculate the % recovery, and comment on the results and the variability in the data provided

Homework Answers

Answer #1

Hopefully this will help you.Thank you.

Know the answer?
Your Answer:

Post as a guest

Your Name:

What's your source?

Earn Coins

Coins can be redeemed for fabulous gifts.

Not the answer you're looking for?
Ask your own homework help question
Similar Questions
ADVERTISEMENT
Need Online Homework Help?

Get Answers For Free
Most questions answered within 1 hours.

Ask a Question
ADVERTISEMENT