Calibration data for a bromide ion-selective electrode (ISE) was collected and recorded below. The potential of the ISE was measured against a saturated calomel electrode (SCE). All solutions were buffered at a pH of 7.56. A linear calibration curve can be constructed from this data as a plot of potential (in mV) vs. pBr. Determine the slope and y-intercept of such a plot of the calibration data given.
[Br-](M) Potential (mV)
0.300 -21.2
0.0300 36.7
0.00300 96.0
0.000300 154.7
0.0000300 225.6
a. slope=
b. y-intercept=
Two unknown solutions were tested for [Br–] using the same bromide ISE. The potential reading for unknown 1 was -6.8 mV and the potential reading for unknown 2 was 172.6 mV. Using the calibration curve you constructed above, determine the concentration of Br– in each unknown solution.
c. Unknown 1 [Br-]=
d. Unknown 2 [Br-]=
If we plot log[Br-] vs potential a linear graph will be obtained for which:
Slope = -57.9 and Intercept = -51.42
Using the value of slope and intercept, the linear equation can be constructed.
y = -57.9x - 51.42
Where y = potential in mV, x =log[Br-]
a. Slope = -57.9
b. y-intercept = -51.42
c. Calculation for unknown 1:
y = -6.8 mV
Substituting the value of y , calculate x.
-6.8 = -57.9x - 51.42
x = -0.77
log[Br-] = -0.77
[Br-] = 0.17 M
d. Calculation for unknown 2:
y = -172.6 mV
Substituting the value of y , calculate x.
-172.6 = -57.9x - 51.42
x = -2.09
log[Br-] = -2.09
[Br-] = 0.008 M