What would be the effect if the cover of the sample compartment is left open while setting 100%T and 0%T?
Select one:
a. 0%T level would be high but 100% T level would be correct
b. 0%T level would be correct but 100% T level would be high
c. No effect
d. Both 0% and 100% levels would be incorrect
Answer: d. Both 0% and 100% levels would be incorrect
Make sure that the sample compartment is empty and the cover is closed, then adjust the Power Switch/Zero Control until the display reads 0%T. Insert the glassware filled with distilled water into the sample compartment and use the Transmittance/Absorbance Control to set the meter or display to 100%T.
I have spectronic user guide and at the troubleshooting part they described above info as shown in below image.
You may also freely access that manual at
http://fg.cns.utexas.edu/fg/course_notebook_appendix_ii_files/Spectronic_20_D.pdf
see page number 39
Hope this helped you!
Thank You So Much! Please Rate this answer as you wish.("Thumbs Up")