Draw and label a box diagram of a spectrophotometer that could use for a UV/Vis absorption experiemt
A diagram of the components of a typical spectrometer are shown in the following diagram. The functioning of this instrument is relatively straightforward. A beam of light from a visible and/or UV light source (colored red) is separated into its component wavelengths by a prism or diffraction grating. Each monochromatic (single wavelength) beam in turn is split into two equal intensity beams by a half-mirrored device. One beam, the sample beam (colored magenta), passes through a small transparent container (cuvette) containing a solution of the compound being studied in a transparent solvent. The other beam, the reference (colored blue), passes through an identical cuvette containing only the solvent. The intensities of these light beams are then measured by electronic detectors and compared. The intensity of the reference beam, which should have suffered little or no light absorption, is defined as I0. The intensity of the sample beam is defined as I. Over a short period of time, the spectrometer automatically scans all the component wavelengths in the manner described. The ultraviolet (UV) region scanned is normally from 200 to 400 nm, and the visible portion is from 400 to 800 nm.
If the sample compound does not absorb light of of a given
wavelength, I = I0.
However, if the sample compound absorbs light then
I is less than I0, and
this difference may be plotted on a graph versus wavelength, as
shown on the right. Absorption may be presented as
transmittance (T =
I/I0) or
absorbance (A= log
I0/I). If no
absorption has occurred, T = 1.0 and A= 0. Most
spectrometers display absorbance on the vertical axis, and the
commonly observed range is from 0 (100% transmittance) to 2 (1%
transmittance). The wavelength of maximum absorbance is a
characteristic value, designated as
?max.
|